Claimed vs Measured: Panel Irradiance Under a Spectrometer

Updated August 2026 · every figure below is parsed from the same database records that power the device pages

Across 44 red light therapy panels where a headline irradiance figure can be checked against a spectrometer reading of the same panel, the headline averages 2.11 times the spectrometer figure (median 2.18, range 1.36 to 2.60). Most of that gap is instrumentation rather than fabrication: a solar power meter reads roughly double what a laboratory spectrometer records on the same LED panel, and the brands transparent enough to publish both readings show the same roughly 2x spread as the brands measured independently.

What counts as a checkable pair

The database records 44 pairs of two kinds. Brand dual disclosures (38 panels): the brand itself publishes a solar meter figure and a spectrometer figure for the same panel, usually both at 6 inches. Publishing both is a transparency practice, and the solar figure is the one comparable to competitors' headline numbers. Independent measurements (6 panels): a third party put a spectrometer on the panel and the database records the matching brand claim; the independent figures here come from the published testing of Light Therapy Insiders and, for one panel, Outliyr. Pairs only qualify when both numbers and their measurement context are recorded in the database; nothing is estimated to fill a row.

The full table

BrandModel or lineClaimed mW/cm²Claimed methodSpectrometer mW/cm²Measured sourceRatio
BlockBlueLightPowerPanel TABLETOP140solar meter, 6 in (brand-published)58brand dual disclosure: spectrometer, 6 in2.41
BlockBlueLightPowerPanel MID158solar meter, 6 in (brand-published)61brand dual disclosure: spectrometer, 6 in2.59
BlockBlueLightPowerPanel PRO160solar meter, 6 in (brand-published)68brand dual disclosure: spectrometer, 6 in2.35
BlockBlueLightPowerPanel MAX162solar meter, 6 in (brand-published)82brand dual disclosure: spectrometer, 6 in1.98
BlockBlueLightPowerPanel MEGA167solar meter, 6 in (brand-published)89brand dual disclosure: spectrometer, 6 in1.88
BlockBlueLightPowerPanel ELITE XL189solar meter, 6 in (brand-published)90brand dual disclosure: spectrometer, 6 in2.10
BlockBlueLightULTIMATE XXL189solar meter, 6 in (brand-published)90brand dual disclosure: spectrometer, 6 in2.10
HoogaULTRA360132brand claim at 6 in, instrument not stated74Light Therapy Insiders: spectrometer, 6 in1.78
HoogaULTRA1500165brand claim at 6 in, instrument not stated81.7Light Therapy Insiders: spectrometer avg, 6 in2.02
Mito Red LightMitoPRO 300X130solar meter, 6 in (brand-published)56brand dual disclosure: spectrometer, 6 in2.32
Mito Red LightMitoPRO 750X160solar meter, 6 in (brand-published)70brand dual disclosure: spectrometer, 6 in2.29
Mito Red LightMitoPRO 1500X160solar meter, 6 in (brand-published)73brand dual disclosure: spectrometer, 6 in2.19
Mito Red LightMitoPRO MegaX130solar meter, 6 in (brand-published)59brand dual disclosure: spectrometer, 6 in2.20
Mito Red LightMitoADAPT MIN 4.0125solar meter, 6 in (brand-published)56brand dual disclosure: spectrometer, 6 in2.23
Mito Red LightMitoADAPT MID 4.0130solar meter, 6 in (brand-published)58brand dual disclosure: spectrometer, 6 in2.24
Mito Red LightMitoADAPT MAX 4.0140solar meter, 6 in (brand-published)60brand dual disclosure: spectrometer, 6 in2.33
Mito Red LightMitoPRO 300+130solar meter, 6 in (brand-published)60brand dual disclosure: spectrometer, 6 in2.17
Mito Red LightMitoPRO 1000+165solar meter, 6 in (brand-published)66brand dual disclosure: spectrometer, 6 in2.50
Mito Red LightMitoPRO 1500+170solar meter, 6 in (brand-published)68brand dual disclosure: spectrometer, 6 in2.50
Mito Red LightMitoMIN 2.0115solar meter, 6 in (brand-published)52brand dual disclosure: spectrometer, 6 in2.21
Mito Red LightMitoMID 2.0150solar meter, 6 in (brand-published)68brand dual disclosure: spectrometer, 6 in2.21
Mito Red LightMitoMAX 2.0155solar meter, 6 in (brand-published)70brand dual disclosure: spectrometer, 6 in2.21
Mito Red LightMitoMEGA 2.0165solar meter, 6 in (brand-published)74brand dual disclosure: spectrometer, 6 in2.23
Mito Red LightMitoPRO Mobile100solar meter, 3 in (brand-published)40brand dual disclosure: spectroradiometer, 3 in2.50
Rouge CareRouge Tabletop G4164solar meter, 6 in (brand-published)79brand dual disclosure: spectrometer, 6 in2.08
Rouge CareRouge Pro G4164solar meter, 6 in (brand-published)90brand dual disclosure: spectrometer, 6 in1.82
Rouge CareRouge Max G4164solar meter, 6 in (brand-published)81brand dual disclosure: spectrometer, 6 in2.02
Rouge CareRouge Ultimate G4164solar meter, 6 in (brand-published)85brand dual disclosure: spectrometer, 6 in1.93
Rojo Light TherapyRefine 180195solar-style comparison figure, 6 in (brand-published)75brand dual disclosure: spectrometer (OHSP350IR), 6 in2.60
Rojo Light TherapyRefine 360195solar-style comparison figure, 6 in (brand-published)75brand dual disclosure: spectrometer (OHSP350IR), 6 in2.60
Rojo Light TherapyRefine 900185solar-style comparison figure, 6 in (brand-published)73brand dual disclosure: spectrometer (OHSP350IR), 6 in2.53
Rojo Light TherapyRefine 1800195solar-style comparison figure, 6 in (brand-published)75brand dual disclosure: spectrometer (OHSP350IR), 6 in2.60
Rojo Light TherapyRefine 3600195solar-style comparison figure, 6 in (brand-published)75brand dual disclosure: spectrometer (OHSP350IR), 6 in2.60
MITO LIGHT (EU)Starter 5.0130solar meter, 15 cm (brand-published)74brand dual disclosure: spectrometer, 15 cm1.76
MITO LIGHT (EU)Expert 5.0140solar meter, 15 cm (brand-published)80brand dual disclosure: spectrometer, 15 cm1.75
PlatinumLEDBioMax 900 (classic)185historical brand claim, 6 in90.3Light Therapy Insiders: spectrometer avg, 6 in2.05
JoovvSolo 3.0100brand claim (a floor), distance not stated72.4Light Therapy Insiders: spectrometer peak1.38
Vital Red LightVital Elite150brand claim 150-210, distance not stated (low end used)78.3Outliyr: spectrometer, 12 in1.92
LightpathLEDDiesel Gen376brand figure at 12 in, method not named56Light Therapy Insiders: spectrometer, 12 in1.36
RLT HomeTotal Spectrum MINI170solar meter, 8 in (brand-published)98.7brand dual disclosure: spectrometer, 6 in1.72
RLT HomeTotal Spectrum COMPACT172solar meter, 8 in (brand-published)99brand dual disclosure: spectrometer, 6 in1.74
RLT HomeTotal Spectrum MAX172solar meter, 8 in (brand-published)100.53brand dual disclosure: spectrometer, 6 in1.71
RLT HomeTotal Spectrum ULTRA174solar meter, 8 in (brand-published)119.28brand dual disclosure: spectrometer, 6 in1.46
RLT HomeTotal Spectrum ELITE179solar meter, 8 in (brand-published)121.62brand dual disclosure: spectrometer, 6 in1.47

Several brand figures are published as floors (a "greater than" number); the floor value is used, which if anything understates the ratio.

How to read this table

Dual disclosure is a credit, not an accusation. A brand listed here because it publishes both readings is being more honest than a brand that publishes one big unlabeled number. The rows measure the gap between instruments, and that gap is exactly why an unlabeled figure deserves skepticism: the marketing convention this table documents is to headline the higher reading.

Peak vs average matters. A spectrometer reading can be reported as the peak at the panel's center or as a multi-point average, and the two differ 15 to 20 percent on the same panel. Where the database records both, the average is used, except for the Joovv row, where the independent peak (the reading most favorable to the brand) is compared against the brand's floor claim.

Distance changes everything. Irradiance falls off steeply with distance, so a claim with no stated distance cannot be cleanly separated from a distance mismatch; the Joovv and Vital Red Light rows carry that caveat. The RLT Home rows compare a solar reading at 8 inches with a spectrometer reading at 6 inches, so the greater solar distance means those ratios understate the pure instrument gap.

Independent tests corroborate the dual disclosures. Where Light Therapy Insiders has also tested a dual-disclosing brand's panel, its spectrometer result lands close to the brand's own published spectrometer figure: the PowerPanel MEGA measured 81 average against a published 89, and the Refine 900 measured 73 average against a published 73. The spectrometer column, in other words, is the credible one.

Limitations

This is 44 panels out of a 188-device database, so it is a sample defined by disclosure, not a random draw. The measurements come from different instruments, operators, and protocols; a handful of pairs compare different distances, as flagged above; several claims are floors; and a solar meter figure is a legitimate reading from a different instrument, not a fabricated number. We do not operate a test lab: every figure here is compiled from published sources recorded in the database, per the methodology.

Reuse this data

This dataset is free to reuse with attribution to RedLightPanelIndex.com. Download the CSV (columns: brand, model, claimed mW/cm², claimed method, measured mW/cm², measured source, ratio).

Independent measurements attributed inline: Light Therapy Insiders and Outliyr, as recorded in the database. Brand figures are from each brand's published product pages, verified per the methodology. See also: why solar meters read roughly double and the dose calculator.